Product Description
Item/Model
BSC-1000ⅡB2
BSC-1300ⅡB2
BSC-1600ⅡB2
Cleanliness Level
HEPA:ISO Class4(Class10)
Supply air velocity
0.33m/s±0.025
Inflow air velocity
0.53m/s±0.025
Noise
≤67dB
Power/Maximum power consumption
AC,220V/50HZ / 1KW
AC,220V/50HZ / 1.5KW
AC,220V/50HZ / 1.5KW
Biosecurity personnel protection
All impact sampler samples CFU≤10, Slit sampler CFU≤5
Product protection
All petri dish samples CFU≤5
Cross-contamination protection
All petri dish samples CFU≤2
Dimension of working area
1000x520x640 mm
1300x520x640 mm
1600x520x640 mm
Overall dimension
1165x760x2200 mm
1465x760x2200 mm
1765x760x2200 mm
Illumination
≥900LX
≥900LX
≥900LX
Weight
240kg
260kg
280kg
Number of Operator
1
1 - 2
2 - 3
Draft direction
Top
1.Type & Airflow:Class II Biosafety cabinet, 100% efflux airflow mode.
2.Standards & Certification:Complies with China’s Pharmaceutical Industry Standard 《YY 0569 - 2011》. Has obtained SFDA medical device registration certificate and production license.
3.Structure & Materials:Outer box: High - quality cold - rolled steel plate, electrostatic - sprayed. Operation area (three sides): Integrated stainless steel structure. Internal cleanable parts: 10mm large rounded corners for easy cleaning. Front window glass door: 6mm thick safety tempered glass.
4.Filtration & Fan:Two HEPA filters; efficiency: ULPA@0.12μm 99.9995%, filter membrane (borosilicate glass fiber, no partition). High - quality fan, auto - adjustable wind speed, low noise.
5.Sensors & Control:High - sensitivity, high - precision breeze speed sensor (non - pressure difference type) for real - time wind speed monitoring. Independently developed control software (national software copyright certificate). LCD screen: Displays temp, airflow velocity, etc.; allows parameter adjustment, password - protected to prevent misoperation.
6.Alarms & Protection:Filter alarm (real - time barcode display of remaining life, sound/light prompt for replacement). Height limit/low limit alarm for moving glass door (sound/light reminder). Unique negative pressure anti - leakage design (working area surrounded by negative pressure, zero leakage). Chain protection: Front window opening → germicidal lamp off, lamp/fan auto - start.
7.UV & Other Features:Front window: Integrated sliding door (arbitrary positioning), UV lamp operable when fully closed. UV lamp: Reservation & timing functions. Spare socket included.
Technical Parameters
Name: Anna Zhao
Mobile:+86 15324910590
Tel:+86 15324910590
Whatsapp:+86-13526736193
Email:Henanshuyu123456@outlook.com
Add:Xiyue City, No. 33 Jinshuo Road, Hi-Tech Industrial Development Zone, Zhengzhou City, Henan Province, China