Specifications
Single-channel Adjustable Volume Pipettes
Volume Range
Increment
Test Volume
Error limits in accordance with ISO8655-2
Systematic Error
Random Error
μL
μL
μL
μL
%
μL
%
0.1–2.5
0.05
2.5
±0.0625
±2.50
±0.05
±2.00
1.25
±0.0375
±3.00
±0.0375
±3.00
0.25
±0.03
±12.00
±0.015
±6.00
0.5–10
0.1
10
±0.1
±1.00
±0.08
±0.80
5
±0.075
±1.50
±0.075
±1.50
1
±0.025
±2.50
±0.015
±1.50
2–20
0.5
20
±0.18
±0.90
±0.08
±0.40
10
±0.12
±1.20
±0.04
±1.00
2
±0.06
±3.00
±0.04
±2.00
5–50
0.5
50
±0.3
±0.60
±0.15
±0.30
25
±0.225
±0.90
±0.15
±0.60
5
±0.1
±2.00
±0.1
±2.00
10–100
1
100
±0.8
±0.80
±0.15
±0.15
50
±0.5
±1.00
±0.2
±0.40
10
±0.3
±3.00
±0.15
±1.50
20–200
1
200
±1.2
±0.60
±0.3
±0.15
100
±0.8
±0.80
±0.3
±0.30
20
±0.6
±3.00
±0.2
±1.00
50–200
1
200
±1.2
±0.60
±0.3
±0.15
100
±0.8
±0.80
±0.3
±0.30
50
±0.5
±1.00
±0.2
±0.40
100–1000
5
1000
±6
±0.60
±2
±0.20
500
±3.5
±0.70
±1.25
±0.25
100
±2
±2.00
±0.7
±0.70
200–1000
5
1000
±6
±0.60
±2
±0.20
500
±3.5
±0.70
±1.25
±0.25
200
±1.8
±0.90
±0.6
±0.30
1000–5000
50
5000
±25
±0.50
±7.5
±0.15
2500
±15
±0.60
±7.5
±0.30
1000
±7
±0.70
±3
±0.30
2000–10000Do not apply to MicroPette plus
100
10000
±60
±0.60
±20
±0.20
5000
±30
±1.20
±15
±0.30
2000
±60
±3.00
±12
±0.60
Fixed Volume Pipettes
Volume Range
Increment
Test Volume
Systematic Error
Random Error
μL
--
μL
μL
%
μL
%
5
--
5
±0.065
±1.3
±0.06
±1.2
10
--
10
±0.08
±0.8
±0.08
±0.8
20
--
20
±0.12
±0.6
±0.1
±0.5
25
--
25
±0.125
±0.5
±0.075
±0.3
50
--
50
±0.25
±0.5
±0.15
±0.3
100
--
100
±0.5
±0.5
±0.3
±0.3
200
--
200
±0.8
±0.4
±0.4
±0.2
250
--
250
±1.0
±0.4
±0.5
±0.2
500
--
500
±1.5
±0.3
±1.0
±0.2
1000
--
1000
±3.0
±0.3
±2.0
±0.2
2000
--
2000
±6.0
±0.3
±3.0
±0.15
5000
--
5000
±15
±0.3
±7.5
±0.15
8-channel Adjustable Volume Pipettes
Volume Range(μL)
Increment(μL)
Test Volume(μL)
Systematic Error(μL)
Systematic Error(%)
Random Error(μL)
Random Error(%)
0.5–10
0.1
10
±0.15
±1.50
±0.15
±1.50
5
±0.125
±2.50
±0.125
±2.50
1
±0.04
±4.00
±0.04
±4.00
5–50
0.5
50
±0.5
±1.00
±0.25
±0.50
25
±0.375
±1.50
±0.25
±1.00
5
±0.15
±3.00
±0.1
±2.00
50–300
5
300
±2.1
±0.70
±0.75
±0.25
150
±1.5
±1.00
±0.75
±0.50
50
±0.75
±1.50
±0.4
±0.80
12-channel Adjustable Volume Pipettes
Volume Range(μL)
Increment(μL)
Test Volume(μL)
Systematic Error(μL)
Systematic Error(%)
Random Error(μL)
Random Error(%)
0.5–10
0.1
10
±0.15
±1.50
±0.15
±1.50
5
±0.125
±2.50
±0.125
±2.50
1
±0.04
±4.00
±0.04
±4.00
5–50
0.5
50
±0.5
±1.00
±0.25
±0.50
25
±0.375
±1.50
±0.25
±1.00
5
±0.15
±3.00
±0.1
±2.00
50–300
5
300
±2.1
±0.70
±0.75
±0.25
150
±1.5
±1.00
±0.75
±0.50
50
±0.75
±1.50
±0.4
±0.80